Eddy current nulled constrained optimization of isotropic diffusion encoding gradient waveforms
Abstract
PURPOSE: Isotropic diffusion encoding efficiently encodes additional microstructural information in combination with conventional linear diffusion encoding. However, the gradient-intensive isotropic diffusion waveforms generate significant eddy currents, which cause image distortions. The purpose of this study is to present a method for designing isotropic diffusion encoding waveforms with intrinsic eddy current nulling.
METHODS: Eddy current nulled gradient waveforms were designed using a constrained optimization waveform for a 3T GE Premier MRI system. Encoding waveforms were designed for a variety of eddy current null times and sequence timings to evaluate the achievable b-value. Waveforms were also tested with both eddy current nulling and concomitant field compensation. Distortion reduction was tested in both phantoms and the in vivo human brain.
RESULTS: The feasibility of isotropic diffusion encoding with intrinsic correction of eddy current distortion and signal bias from concomitant fields was demonstrated. The constrained optimization algorithm produced gradient waveforms with the specified eddy current null times. The reduction in the achievable diffusion weighting was dependent on the number of eddy current null times. A reduction in the eddy current-induced image distortions was observed in both phantoms and in vivo human subjects.
CONCLUSION: The proposed framework allows the design of isotropic diffusion-encoding sequences with reduced image distortion.
Read the paper
Related studies
- Learning from First-Generation Qualitative Approaches in the IS Discipline: An Evolutionary View and Some Implications for Authors and Evaluators (PART 1/2)
- Commentary on the clinical and preclinical dosage limits of interstitially administered magnetic fluids for therapeutic hyperthermia based on current practice and efficacy models
- Prosthesis: New Technological Opportunities and Innovative Biomedical Devices
- The history and future of neural modeling for cochlear implants
- Ryanodine receptors: waking up from refractoriness
- Novel Transgenic Rabbit Model Sheds Light on the Puzzling Role of Matrix Metalloproteinase-12 in Atherosclerosis
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.