Variability of renal apparent diffusion coefficients: limitations of the monoexponential model for diffusion quantification
Abstract
PURPOSE: To investigate whether variability in reported renal apparent diffusion coefficient (ADC) values in literature can be explained by the use of different diffusion weightings (b values) and the use of a monoexponential model to calculate ADC.
MATERIALS AND METHODS: This prospective study was approved by institutional review board and was HIPAA-compliant, and all subjects gave written informed consent. Diffusion-weighted (DW) imaging of the kidneys was performed in three healthy volunteers to generate reference diffusion decay curves. In a literature meta-analysis, the authors resampled the reference curves at the various b values used in 19 published studies of normal kidneys (reported ADC = [2.0-4.1] x 10(-3) mm(2) / sec for cortex and [1.9-5.1] x 10(-3) mm(2) / sec for medulla) and then fitted the resampled signals by monoexponential model to produce "predicted" ADC. Correlation plots were used to compare the predicted ADC values with the published values obtained with the same b values.
RESULTS: Significant correlation was found between the reported and predicted ADC values for whole renal parenchyma (R(2) = 0.50, P = .002), cortex (R(2) = 0.87, P = .0002), and medulla (R(2) = 0.61, P = .0129), indicating that most of the variability in reported ADC values arises from limitations of a monoexponential model and use of different b values.
CONCLUSION: The use of a monoexponential function for DW imaging analysis and variably sampled diffusion weighting plays a…
How strong is the evidence?
Evidence Hub reads this as a meta-analysis, done in people. Its evidence level is 1 of 5 (Highest · pooled evidence). Pooled evidence from several studies sits at the top of the evidence pyramid.
Read the paper
Related studies
- Learning from First-Generation Qualitative Approaches in the IS Discipline: An Evolutionary View and Some Implications for Authors and Evaluators (PART 1/2)
- Commentary on the clinical and preclinical dosage limits of interstitially administered magnetic fluids for therapeutic hyperthermia based on current practice and efficacy models
- Prosthesis: New Technological Opportunities and Innovative Biomedical Devices
- The history and future of neural modeling for cochlear implants
- Ryanodine receptors: waking up from refractoriness
- Novel Transgenic Rabbit Model Sheds Light on the Puzzling Role of Matrix Metalloproteinase-12 in Atherosclerosis
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.