Extremely efficient and deterministic approach to generating optimal ordering of diffusion MRI measurements
Abstract
PURPOSE: Diffusion MRI measurements are typically acquired sequentially with unit gradient directions that are distributed uniformly on the unit sphere. The ordering of the gradient directions has significant effect on the quality of dMRI-derived quantities. Even though several methods have been proposed to generate optimal orderings of gradient directions, these methods are not widely used in clinical studies because of the two major problems. The first problem is that the existing methods for generating highly uniform and antipodally symmetric gradient directions are inefficient. The second problem is that the existing methods for generating optimal orderings of gradient directions are also highly inefficient. In this work, the authors propose two extremely efficient and deterministic methods to solve these two problems.
METHODS: The method for generating nearly uniform point set on the unit sphere (with antipodal symmetry) is based upon the notion that the spacing between two consecutive points on the same latitude should be equal to the spacing between two consecutive latitudes. The method for generating optimal ordering of diffusion gradient directions is based on the idea that each subset of incremental sample size, which is derived from the prescribed and full set of gradient directions, must be as uniform as possible in terms of the modified electrostatic energy designed for antipodally symmetric point set.
RESULTS: The proposed method outperformed the…
Read the paper
Related studies
- Learning from First-Generation Qualitative Approaches in the IS Discipline: An Evolutionary View and Some Implications for Authors and Evaluators (PART 1/2)
- Commentary on the clinical and preclinical dosage limits of interstitially administered magnetic fluids for therapeutic hyperthermia based on current practice and efficacy models
- Prosthesis: New Technological Opportunities and Innovative Biomedical Devices
- The history and future of neural modeling for cochlear implants
- Ryanodine receptors: waking up from refractoriness
- Novel Transgenic Rabbit Model Sheds Light on the Puzzling Role of Matrix Metalloproteinase-12 in Atherosclerosis
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.