Microcalorimeter EDS: Benefits and Drawbacks

David A. Wollman, Dale E. Newbury, Sae Woo Nam, G. C. Hilton, K. D. Irwin, David A. Rudman, Steven W. Deiker, Norman F. Bergren et al.

Microscopy and Microanalysis, 2000

1 citation

Abstract

Abstract The commercial introduction of high-count-rate, near-room-temperature silicon drift detectors (presently available) and high-energy-resolution cryogenic microcalorimeters (forthcoming) is an exciting development in x-ray microanalysis, in which detector choices and capabilities have been essentially stable for many years. Both of these new energy-dispersive detectors promise improved capabilities for specific applications, e.g., faster EDS mapping (silicon drift detectors) and nanoscale particle analysis (microcalorimeters). In this paper, we briefly examine some of the important benefits and drawbacks of microcalorimeter EDS (μcal EDS) for x-ray microanalysis. The primary benefit of μcal EDS over conventional semiconductor EDS is the factor of ∼ 20 improvement in energy resolution (∼ 4 eV, real-time analog signal processing), as shown in Figure 1.