Abstract
This paper describes a rugged and precise analytical method developed and validated for determining the particle size distribution of Empagliflozin drug substance using a wet dispersion technique on the Malvern Mastersizer 3000. The method was optimized to achieve uniform dispersion and reproducible measurements of particle size parameters (D10, D50, D90). Validation of the method was carried out in accordance with ICH Q2(R2) guideline and demonstrated satisfactory results for precision, intermediate precision, and robustness. All validation parameters were found within the predefined acceptance criteria, confirming the reliability of the procedure. Hence, the developed method is suitable for particle size analysis of Empagliflozin active pharmaceutical ingredient (API).