Discordance of longitudinal changes in bone density between densitometers
Abstract
UNLABELLED: This study examined the concordance in BMD measurement and longitudinal change in BMD between the GE Lunar Prodigy and GE Lunar DPX. Even though a high concordance between the densitometers was observed on a single measurement occasion, a significant discordance in longitudinal changes in BMD was observed.
INTRODUCTION: Measurement of bone mineral density (BMD) by dual-energy X-ray absorptiometry (DXA) technology plays an important role in the diagnosis and management of osteoporosis. The present study examined the concordance in BMD measurement and longitudinal change in BMD between GE Lunar Prodigy and DPX.
METHODS: BMD at the lumbar spine and femoral neck was measured in 135 individuals (47 men and 88 women, mean age 73+/-9 years) using both GE Lunar DPX and Prodigy densitometers at baseline. In this group, 56 individuals (22 men and 34 women) had repeated BMD measurements using the DPX and Prodigy during a subsequent follow-up visit (average duration: 2.2 years).
RESULTS: For a single BMD measurement, the coefficient of concordance between the Prodigy and DPX was greater than 0.98 at the lumbar spine and 0.96 at the femoral neck, with the slope of linear regression being approximately 1.0. During the period of follow-up, the lumbar spine BMD decreased by -0.5% (S.D. 1.8%) when measured by DPX, which was significantly different (p=0.002) from the change measured by Prodigy (mean change=0, S.D. 2.0%). However, there was no significant difference (p=0.95) in the…
How strong is the evidence?
Evidence Hub reads this as a cohort study, done in people. Its evidence level is 3 of 5 (Moderate · trial / cohort / secondary analysis). Trials without a large randomized design, cohorts and secondary analyses give moderate evidence. Reported size: 135 participants or records.
Read the paper
Related studies
- Development of a crushable foam model for human trabecular bone
- Data-driven motion compensation in HR-pQCT.
- Modeling of structure, quality, and function in the orthodontic patient
- Exploring the mechanism of Heidihuang Pill in the treatment of osteoporosis based on network pharmacology, molecular docking, and experimental validation
- Effect of fabric on the accuracy of computed tomography-based finite element analyses of the vertebra
- Use of interventions involving virtual reality tasks during the climacteric: a systematic review
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.