IROC phantoms accurately detect MLC delivery errors

Sharbacha S Edward, J. Pollard-Larkin, Peter A. Balter, Rebecca M. Howell, Christine B. Peterson, S. Kry

Journal of Applied Clinical Medical Physics, 2025

2 citations

Abstract

Abstract Purpose We evaluated the impact of random and whole‐bank multileaf collimator (MLC) delivery errors on dosimetric delivery accuracy in the Imaging and Radiation Oncology Core (IROC) phantom audits, as well as differences in delivery accuracy between the IROC phantom prescription and typical clinical fraction sizes. Methods and Materials Plans were created for the IROC IMRT head and neck (H&N) and SBRT spine phantoms. MLC leaf errors were introduced into the plans: random shifts between −2 and 2 mm, and whole bank shifts of 0.5, 1, and 2 mm. Plans were recalculated and delivered on a Varian Truebeam, and the log files were analyzed using Mobius Fx software. A second study examined the impact of fraction size on MLC position accuracy and corresponding dose delivery accuracy. The standard IROC phantom prescriptions (∼6 Gy) were scaled to the extremes of 2 Gy for H&N and 27 Gy for spine. All plans (original and scaled) were delivered on a Varian Truebeam and 21EX machine. Results Random MLC positioning errors produced small average dose deviations in the PTV of up to −2.8% for H&N and 0.7% for spine. Whole‐bank MLC shifts resulted in larger average PTV dose deviations up to 8% for H&N and 7.1% for spine. The Varian 21EX irradiations had greater MLC root mean square (RMS) error than Truebeam plans. Plans with smaller prescriptions (and faster leaf motion) had greater MLC RMS errors, but plan accuracy was not affected dosimetrically – all results remained within 1%…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free