Assessing focal spot alignment in clinical linear accelerators: a comprehensive evaluation with triplet phantoms

H. L. Riis, Kenni H Engstrøm, Luke A Slama, J. Dass, M. Ebert, P. Rowshanfarzad

Physical and Engineering Sciences in Medicine, 2024

1 citation

Abstract

A fundamental parameter to evaluate the beam delivery precision and stability on a clinical linear accelerator (linac) is the focal spot position (FSP) measured relative to the collimator axis of the radiation head. The aims of this work were to evaluate comprehensive data on FSP acquired on linacs in clinical use and to establish the ability of alternative phantoms to detect effects on patient plan delivery related to FSP. FSP measurements were conducted using a rigid phantom holding two ball-bearings at two different distances from the radiation source. Images of these ball-bearings were acquired using the electronic portal imaging device (EPID) integrated with each linac. Machine QA was assessed using a radiation head-mounted PTW STARCHECK phantom. Patient plan QA was investigated using the SNC ArcCHECK phantom positioned on the treatment couch, irradiated with VMAT plans across a complete 360° gantry rotation and three X-ray energies. This study covered eight Elekta linacs, including those with 6 MV, 18 MV, and 6 MV flattening-filter-free (FFF) beams. The largest range in the FSP was found for 6 MV FFF. The FSP of one linac, retrofitted with 6 MV FFF, displayed substantial differences in FSP compared to 6 MV FFF beams on other linacs, which all had FSP ranges less than 0.50 mm and 0.25 mm in the lateral and longitudinal directions, respectively. The PTW STARCHECK phantom proved effective in characterising the FSP, while the SNC ArcCHECK measurements could not discern…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free