Image-Guided Radiotherapy Using a Modified Industrial Micro-CT for Preclinical Applications
Abstract
Purpose/Objective Although radiotherapy is a key component of cancer treatment, its implementation into pre-clinical in vivo models with relatively small target volumes is frequently omitted either due to technical complexity or expected side effects hampering long-term observational studies. We here demonstrate how an affordable industrial micro-CT can be converted into a small animal IGRT device at very low costs. We also demonstrate the proof of principle for the case of partial brain irradiation of mice carrying orthotopic glioblastoma implants. Methods/Materials A commercially available micro-CT originally designed for non-destructive material analysis was used. It consists of a CNC manipulator, a transmission X-ray tube (10–160 kV) and a flat-panel detector, which was used together with custom-made steel collimators (1–5 mm aperture size). For radiation field characterization, an ionization chamber, water-equivalent slab phantoms and radiochromic films were used. A treatment planning tool was implemented using a C++ application. For proof of principle, NOD/SCID/γc−/− mice were orthotopically implanted with U87MG high-grade glioma cells and irradiated using the novel setup. Results The overall symmetry of the radiation field at 150 kV was 1.04±0.02%. The flatness was 4.99±0.63% and the penumbra widths were between 0.14 mm and 0.51 mm. The full width at half maximum (FWHM) ranged from 1.97 to 9.99 mm depending on the collimator aperture size. The dose depth curve along…
Read the paper
Related studies
- Lathyrism-induced alterations in collagen cross-links influence the mechanical properties of bone material without affecting the mineral
- High-Resolution SPECT-CT/MR Molecular Imaging of Angiogenesis in the Vx2 Model
- Severely impaired bone material quality in Chihuahua zebrafish resembles classical dominant human osteogenesis imperfecta
- Small-animal SPECT and SPECT/CT: application in cardiovascular research
- Intravital imaging of osteocytes in mouse calvaria using third harmonic generation microscopy
- Techniques and Evaluation From a Cross-Platform Imaging Comparison of Quantitative Ultrasound Parameters in an In Vivo Rodent Fibroadenoma Model
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.