Monte Carlo analysis of beam blocking grid design parameters: Scatter estimation and the importance of electron backscatter

G. Bootsma, L. Ren, Hong Zhang, Jian-Yue Jin, D. Jaffray

Medical Physics (Lancaster), 2018

4 citations

Abstract

PURPOSE: Beam blocking grids provide a simple and direct measurement of the scattered photon signal which degrades image quality in x-ray imaging systems, such as cone-beam CT (CBCT). This study evaluates the scatter estimation accuracy of the beam blocking method to optimize the design parameters of the grid system (e.g., grid thickness, source-to-grid distance (SGD), septa width, air interspace, and grid ratio) using Monte Carlo (MC) simulations.

METHOD: A MC model of a CBCT imaging system with a beam blocking grid in place is made using code based on EGSnrc, with the x-ray tube portion of the simulation including electron backscatter between the anode and cathode. The inclusion of the electron backscatter allowed a more complete model of the contamination signal to be estimated. The contamination signal consists of the off-focal radiation (OFR) and source component scatter (photon scatter in source components such as tube housing, filters, and collimators). The MC model was validated against measurements collected on a bench top imaging system with a grid in place. The MC model was used to simulate 11 different grid design configurations in addition to a case with no grid. For each design a simulated projection with and without a phantom in place was computed. The simulated projections were then used to estimate the scatter and contamination portion of the signal using the signal behind the grid septa. The estimated signals from the grid data were compared to the actual…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free