Clinical implementation and initial experience of real-time motion tracking with jaws and MLC during helical tomotherapy delivery

Guang-Pei Chen, A. Tai, L. Puckett, E. Gore, Sara N Lim, T. Keiper, C. Johnstone, M. Shukla et al.

Practical Radiation Oncology, 2021

29 citations

Abstract

PURPOSE

This work reports the clinical implementation of a real-time motion tracking and correction system using dynamic multileaf collimator (MLC) and jaws during helical tomotherapy delivery (Synchrony on Radixact, Accuray Inc).

MATERIALS AND METHODS

The first clinical Synchrony on Radixact system was recently installed and tested at our institution. Various clinical workflows, including fiducial implantation, CT simulation, treatment planning, delivery quality assurance (DQA), treatment simulation, and delivery, for both fiducial-free and fiducial-based motion tracking methods were developed. Treatment planning and delivery data from initial patients, including dosimetric benefits, real-time target detection, model building, motion tracking accuracy, delivery smoothness, and extra dose from real-time radiographic imaging, were analyzed.

RESULTS

The Synchrony on Radixact system was tested to be within its performance specifications and has been used to treat 10 lung (fiducial-free) and 5 prostate (fiducial-based) cancer patients so far in our clinic. The success of these treatments, especially for fiducial-free tracking, depends on multiple factors including careful selection of the patient, appropriate setting of system parameters, appropriate positioning of the patient and skin markers, and use of treatment simulation. For the lung tumor cases, the difficulties in model building due primarily to the changes of target detectability and/or respiration patterns, were…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free