Assessment of Dosimetric Accuracy and Plan Quality in Prostate SBRT: Retrospective Evaluation of 36.25 Gy in 5 Fractions Using RapidArc on Varian Millennium MLC with MLC Parameter Analysis
Abstract
Introduction Prostate cancer is one of the most commonly diagnosed cancers among men worldwide. Advances in radiotherapy, particularly Stereotactic Body Radiotherapy (SBRT), have enabled ultra-hypofractionated treatment schedules that enhance tumor control while reducing treatment time. This study focuses on evaluating the dosimetric accuracy and plan quality of prostate SBRT using RapidArc technology on a Varian Millennium Multileaf Collimator (MLC) system. Objectives To retrospectively evaluate the dosimetric parameters, plan quality metrics, and multileaf collimator performance in prostate SBRT delivered via RapidArc, and to assess target coverage, organ-at-risk (OAR) constraints, and delivery efficiency. Materials and Methods A total of 24 patients with localized prostate adenocarcinoma received SBRT with a prescribed dose of 36.25 Gy in five fractions. Treatment planning was performed using Eclipse v15.6 with Acuros XB algorithm, utilizing three 6 MV flattening filter-free (FFF) arcs. Planning Target Volume (PTV) coverage, OAR doses, Paddick Conformity Index (PCI), Gradient Index (GI), and Monitor Units (MU) were analyzed. Multileaf collimator motion was evaluated through log files, including leaf speed and position dynamics. Quality assurance was performed using electronic portal imaging devices (EPID) and gamma pass rate evaluation. Results The mean PTV D95 was 35.80 +/- 0.46 Gy, with mean Dmax and Dmean being 39.80 +/- 1.05 Gy and 37.10 +/- 0.40 Gy, respectively.…
Read the paper
Related studies
- Lathyrism-induced alterations in collagen cross-links influence the mechanical properties of bone material without affecting the mineral
- High-Resolution SPECT-CT/MR Molecular Imaging of Angiogenesis in the Vx2 Model
- Severely impaired bone material quality in Chihuahua zebrafish resembles classical dominant human osteogenesis imperfecta
- Small-animal SPECT and SPECT/CT: application in cardiovascular research
- Intravital imaging of osteocytes in mouse calvaria using third harmonic generation microscopy
- Techniques and Evaluation From a Cross-Platform Imaging Comparison of Quantitative Ultrasound Parameters in an In Vivo Rodent Fibroadenoma Model
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.