Scan duration for dopamine transporter neuroimaging with[123I]FP-CIT and a high-sensitivity multiple-pinhole SPECT/CT

G. Thomsen, K. Madsen, Cecilie A. Poulsen, C. Svarer, L. H. Pinborg, G. M. Knudsen, R. de Nijs

EJNMMI Physics, 2026

1 citation

Abstract

Background: Dopamine transporter (DAT) imaging with [123I]FP-CIT (Ioflupane, DaTscan™) is an established diagnostic tool for diagnosing parkinsonism and usually requires 40 min. of scanning. The three-headed high-sensitivity multiple-pinhole SPECT/CT scanner with dedicated brain multi-pinhole collimators (MPH) has in many respects superior performance and potentially enables a reduced scan duration, which can be a major advantage for patients and increases the number of patient studies that can be performed in one day. The aim of this study is to determine the minimal required scan duration that still returns good image quality and diagnostic accuracy in parkinsonism.

Methods: Fifty patients suspected of parkinsonism (age range 52-87 years) and 9 healthy controls (age range 57-66 years) underwent imaging with Mediso AnyScan Trio SPECT/CT and MPH collimators for 40 min. three hours after injection of 185 MBq [123I]FP-CIT. The images were reconstructed using four different acquisition lengths: 10, 20, 30, and 40 min., and subsequently quantified with an automatic image-based analysis. Outcomes of the different acquisition lengths were compared by statistical analysis of the SBR-values and by visual assessment by experienced physicians.

Results: Compared to 40 min. acquisitions and both in patients and healthy controls, 10 min acquisitions resulted in a significantly higher specific binding ratio (SBR) (≥ 0.5) in the striatum, putamen and caudate (p ≤ 0.001). SBR did not differ…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free