Description and dosimetric verification of the PEREGRINE Monte Carlo dose calculation system for photon beams incident on a water phantom
Abstract
PEREGRINE is a three-dimensional Monte Carlo dose calculation system written specifically for radiotherapy. This paper describes the implementation and overall dosimetric accuracy of PEREGRINE physics algorithms, beam model, and beam commissioning procedure. Particle-interaction data, tracking geometries, scoring, variance reduction, and statistical analysis are described. The BEAM code system is used to model the treatment-independent accelerator head, resulting in the identification of primary and scattered photon sources and an electron contaminant source. The magnitude of the electron source is increased to improve agreement with measurements in the buildup region in the largest fields. Published measurements provide an estimate of backscatter on monitor chamber response. Commissioning consists of selecting the electron beam energy, determining the scale factor that defines dose per monitor unit, and describing treatment-dependent beam modifiers. We compare calculations with measurements in a water phantom for open fields, wedges, blocks, and a multileaf collimator for 6 and 18 MV Varian Clinac 2100C photon beams. All calculations are reported as dose per monitor unit. Aside from backscatter estimates, no additional, field-specific normalization is included in comparisons with measurements. Maximum discrepancies were less than either 2% of the maximum dose or 1.2 mm in isodose position for all field sizes and beam modifiers.
Read the paper
Related studies
- Characterization and Analysis of Porosity and Pore Structures
- Practical considerations for conducting ecotoxicity test methods with manufactured nanomaterials: what have we learnt so far?
- Dental Optical Coherence Tomography
- Radiation-shielding concrete: A review of materials, performance, and the impact of radiation on concrete properties
- Testing entanglement of annihilation photons
- Calanus finmarchicus seasonal cycle and diapause in relation to gene expression, physiology, and endogenous clocks
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.