Quantitative Imaging With a Pinhole Cardiac SPECT CZT Camera

Amir Pourmoghaddas

2016

Abstract

This work develops methods to quantify absolute activity measurements with a multi-head pinhole cardiac SPECT camera that uses cadmium-zinc-telluride (CZT) detectors (Discovery NM530c, GE Healthcare).A central component of absolute activity quantification is the correction for Compton scattered photons.A modified Dual Energy Window (DEW) scatter correction (SC) method was developed that compensates for the presence of unscattered photons in the lower-energy window used to measure scatter.The DEW-SC method was validated using phantom experiments.The mean error in absolute activity measurement was 5 ± 4 % when correcting for attenuation, scatter and the partial volume effects.Clinical accuracy was assessed using 99m Tc-tetrofosmin myocardial perfusion images obtained on the NM530c and compared to DEW scatter-corrected images acquired on a conventional SPECT camera for the same patient.DEW-SC images acquired on the NM530c had increased noise but had summed perfusion scores that were not significantly different from those acquired on the conventional SPECT camera.To reduce noise in scatter corrected images and provide a more accurate estimate of scatter, a model-based SC method was developed based on the analytical photon distribution (APD) approach.Using physical phantom experiments and a set of five clinical studies, the accuracy of APD-SC was evaluated and compared to DEW-SC.Images generated using the model-based method agreed well with acquired data.APD-SC images had lower…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free