Electrical characteristics of an electronic control device under a physiologic load: a brief report

Dawes DM, Ho JD, Kroll MW, Miner JR

Pacing and clinical electrophysiology : PACE, 2010

Abstract

BACKGROUND: Law enforcement officers use electronic control devices (ECDs), such as the TASER X26 (TASER International, Inc., Scottsdale, AZ, USA), to control resisting subjects. Some of the debate on the safety of the devices has centered on the electrical characteristics of the devices. The electrical characteristics published by TASER International have historically based on discharges into a 400 Omega resistor. There are no studies that the authors are aware of that report the electrical characteristics under a physiologic load. In this study, we make an initial attempt to determine the electrical characteristics of the TASER X26 during a 5-second exposure in human volunteers.

METHODS: Subjects received an exposure to the dry, bare chest (top probe), and abdomen (bottom probe) with a standard TASER X26 in the probe deployment mode for 5 seconds. There were 10-11 pulse captures during the 5 seconds. Resistance was calculated using the sum of the absolute values of the instantaneous voltage measurements divided by the sum of the absolute values of the instantaneous current measurements (Ohm's Law).

RESULTS: For the eight subjects, the mean spread between top and bottom probes was 12.1 inches (30.7 cm). The mean resistance was 602.3 Omega, with a range of 470.5-691.4 Omega. The resistance decreased slightly over the 5-second discharge with a mean decrease of 8.0%. The mean rectified charge per pulse was 123.0 microC. The mean main phase charge per pulse was 110.5 microC.…

How strong is the evidence?

Evidence Hub reads this as a cohort study, done in people. Its evidence level is 3 of 5 (Moderate · trial / cohort / secondary analysis). Trials without a large randomized design, cohorts and secondary analyses give moderate evidence.

How the five levels of evidence work. The level is assigned automatically from the abstract and can be wrong, so check the paper.

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free