Abstract
BACKGROUND: Remote magnetic catheter navigation (MNS) has been shown to be feasible and safe for radiofrequency catheter ablation of various cardiac arrhythmias. However, its safety in patients with implanted pacemakers or cardioverter-defibrillators has not yet been studied.
OBJECTIVES: This retrospective case series study intends to assess the acute and short-term safety of remote MNS in patients with implanted pacemakers or cardioverter-defibrillators.
METHODS: Between January 2008 and June 2009, a total of 31 patients with implanted pacemakers (n = 5) or cardioverter-defibrillators (n = 26) underwent 32 catheter ablation procedures using the remote MNS. Baseline pacing thresholds, sensed amplitudes, pacing and, if available, shock impedances as well as battery status were measured in all devices before, immediately after, and 1-3 months after catheter ablation.
RESULTS: After ablation, no statistically significant difference in atrial sensing (2.7 ± 1.5 mV vs 3.1 ± 1.9 mV, P = 0.18) and impedance (457 ± 104 Ω vs 449 ± 101 Ω, P = 0.37) were observed. After ablation, no statistically significant difference in right ventricular sensing (10.4 ± 3.8 mV vs 10.9 ± 4.9 mV, P = 0.43) and impedance (535 ± 118 Ω vs 534 ± 120 Ω, P = 0.913) were observed. No changes in pacing threshold could be observed in all but 2 patients with biventricular cardioverter-defibrillators who underwent ventricular tachycardia ablation in lateral wall of left ventricle near the implanted epicardial…
How strong is the evidence?
Evidence Hub reads this as a case series, done in people. Its evidence level is 5 of 5 (Very low · case / narrative / animal / in vitro / in silico). Case reports, narrative reviews, and animal, cell or computer-model studies give the weakest evidence for human health. Reported size: 31 participants or records.
Read the paper
Related studies
- eggNOG 5.0: a hierarchical, functionally and phylogenetically annotated orthology resource based on 5090 organisms and 2502 viruses
- Online Survival Analysis Software to Assess the Prognostic Value of Biomarkers Using Transcriptomic Data in Non-Small-Cell Lung Cancer
- OptiType: precision HLA typing from next-generation sequencing data
- Nicotinic Cholinergic System and COVID-19: In Silico Identification of an Interaction between SARS-CoV-2 and Nicotinic Receptors with Potential Therapeutic Targeting Implications
- Saccharomyces cerevisiae cells have three Omega class glutathione S-transferases acting as 1-Cys thiol transferases
- Hemolysis at low blood flow rates: in-vitro and in-silico evaluation of a centrifugal blood pump
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.