Low-voltage shock impedance measurements: A false sense of security
Abstract
BACKGROUND: Implantable cardioverter defibrillators use low-voltage shock impedance measurements to monitor the lead integrity. However, previous case reports suggest that low-voltage shock impedance measurements may fail to detect insulation breaches that can cause life-threatening electrical short circuits.
METHODS AND RESULTS: We report six cases of insulation breaches in transvenous defibrillation leads that were not obvious during standard interrogations and testing of the lead beforehand. In two cases, an electrical short circuit during commanded shock delivery for internal electrical cardioversion resulted in a total damage of the ICD generator. In one of these cases, commanded shock delivery induced ventricular fibrillation, which required external defibrillation. In two cases, a shock due to ventricular tachycardia was aborted as the shock impedance was less than 20 Ω. However, in both cases the tiny residual shock energy terminated the ventricular tachycardia. In contrast, in one case the residual energy of the aborted shock did not end ventricular fibrillation induced at defibrillator threshold testing. In one case, the ICD indicated an error code for a short circuit condition detected during an adequate shock delivery.
CONCLUSIONS: This case series illustrates that low-voltage shock impedance measurements can fail to detect insulation breaches. These data suggest that in patients without a contraindication, traditional defibrillator threshold testing or high…
How strong is the evidence?
Evidence Hub reads this as a case series, done in people. Its evidence level is 5 of 5 (Very low · case / narrative / animal / in vitro / in silico). Case reports, narrative reviews, and animal, cell or computer-model studies give the weakest evidence for human health.
Read the paper
Related studies
- eggNOG 5.0: a hierarchical, functionally and phylogenetically annotated orthology resource based on 5090 organisms and 2502 viruses
- Online Survival Analysis Software to Assess the Prognostic Value of Biomarkers Using Transcriptomic Data in Non-Small-Cell Lung Cancer
- OptiType: precision HLA typing from next-generation sequencing data
- Nicotinic Cholinergic System and COVID-19: In Silico Identification of an Interaction between SARS-CoV-2 and Nicotinic Receptors with Potential Therapeutic Targeting Implications
- Saccharomyces cerevisiae cells have three Omega class glutathione S-transferases acting as 1-Cys thiol transferases
- Hemolysis at low blood flow rates: in-vitro and in-silico evaluation of a centrifugal blood pump
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.