Usefulness of electroanatomical mapping during transseptal endocardial left ventricular lead implantation
Abstract
AIM: Failure rate to implant left ventricular (LV) lead transvenously is 4-8% in cardiac resynchronization therapy (CRT) patients. Epicardial lead placement is an alternative method and if not applicable case reports and small series showed the feasibility of endocardial LV lead implantation. Electroanatomical mapping might be a useful tool to guide this procedure.
METHODS AND RESULTS: Four patients had undergone endocardial LV lead implantation after unsuccessful transvenous implantation or epicardial LV lead dysfunction using the transseptal approach. Electroanatomical mapping was used to mark the location of the transseptal puncture. This location point guided the mapping catheter from the subclavian access and facilitated positioning of the LV lead at the adjacent latest activation area of the left ventricle detected by activation mapping. Endocardial active fixation LV leads were successfully implanted in all patients with stable electrical parameters immediately after implantation and over a mean follow-up of 18.3 months (lead impedance 520 ± 177 vs. 439 ± 119 Ω and pacing threshold 0.8 ± 0.2 V, 0.5 ms vs. 0.6 ± 0.1 V, 0.5 ms, respectively). Patients were maintained on anticoagulation therapy with a target international normalized ratio of 3.5-4.5 and did not show any thromboembolic, haemorrhagic events, or infection. Echocardiography showed significant improvement of LV systolic function with marked improvement of the functional status.
CONCLUSIONS: Electroanatomical…
How strong is the evidence?
Evidence Hub reads this as a case report, done in people. Its evidence level is 5 of 5 (Very low · case / narrative / animal / in vitro / in silico). Case reports, narrative reviews, and animal, cell or computer-model studies give the weakest evidence for human health.
Read the paper
Related studies
- eggNOG 5.0: a hierarchical, functionally and phylogenetically annotated orthology resource based on 5090 organisms and 2502 viruses
- Online Survival Analysis Software to Assess the Prognostic Value of Biomarkers Using Transcriptomic Data in Non-Small-Cell Lung Cancer
- OptiType: precision HLA typing from next-generation sequencing data
- Nicotinic Cholinergic System and COVID-19: In Silico Identification of an Interaction between SARS-CoV-2 and Nicotinic Receptors with Potential Therapeutic Targeting Implications
- Saccharomyces cerevisiae cells have three Omega class glutathione S-transferases acting as 1-Cys thiol transferases
- Hemolysis at low blood flow rates: in-vitro and in-silico evaluation of a centrifugal blood pump
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.