Delivery System Support Prevents Endocardial Entanglement and Delamination During Left Bundle Branch Area Pacing

Miyazaki Y, Nakajima H, Rimmer M, Hanna P, Hayase JH, Do DH, Shivkumar K, Mori S

Circulation. Arrhythmia and electrophysiology, 2026

Abstract

Background: The mechanisms behind endocardial entanglement and resultant delamination during left bundle branch area pacing, as well as potential solutions, have not been fully demonstrated in the human heart.

Methods: Active fixation of a lumenless screw-in lead into the right ventricular septum was attempted with or without delivery system support using donor human hearts rejected for transplant. Delivery sheath support was ensured through firm catheter support maintained by a fixed clip. Endocardial entanglement, endocardial delamination after unscrewing, and lead penetration were evaluated using both macroscopic and microscopic analyses.

Results: Each of 10 lead deployments with or without support was attempted using 4 donor human hearts. There was a higher incidence of endocardial entanglement and resultant helix-only penetration in the nonsupport group than the support group (70% versus 0%). Unscrewing after the endocardial entanglement consistently revealed extensively delaminated endocardium entwined around the helix. The median endocardial delamination area was larger in the entanglement group than in the nonentanglement group (25.0 mm2 versus 0.9 mm2, P<0.001). In the regions with endocardial entanglement, histological analysis confirmed extensive endocardial delamination with minimal myocardial injury. On the other hand, localized endocardial delamination and deep intramural penetration were observed in the regions without entanglement. Screw-in procedures…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free