Magnetic resonance monitoring of stent deployment: in vitro evaluation of different stent designs and stent delivery systems
Abstract
RATIONALE AND OBJECTIVES: To evaluate MR imaging features of commercially available stents before, during, and after in vitro deployment as a step toward MR-guided stent deployment.
METHODS: Fourteen stents were deployed in a phantom under MR monitoring at 1.5 T by using a gradient-echo sequence. Device visibility was rated on a four-point scale (excellent, fair, poor, not visible).
RESULTS: The Memotherm stent and the rolling membrane (RM) Wallstent showed excellent stent visibility and at least fair scores for artifact-induced narrowing of the stent lumen. Three stents (Palmaz, AVE, Easy Wallstent) showed excellent visibility of the stent but no visible lumen. Five stents (Strecker, Accuflex, Hemobahn, Passager, Sinus) displayed fair visibility. The delivery catheters of four stent systems (Smart, Vascucoil, Symphony, ZA) displayed severe black hole artifacts.
CONCLUSIONS: The imaging features of several stent systems might be suitable for MR-guided intervention. The Memotherm and the Wallstent RM combine good visibility of the stent and the lumen.
Read the paper
Related studies
- A Survey on Knowledge Graphs: Representation, Acquisition, and Applications
- Toxicity of nanomaterials
- The Human Gene Mutation Database: towards a comprehensive repository of inherited mutation data for medical research, genetic diagnosis and next-generation sequencing studies
- Halogen bonding (X‐bonding): A biological perspective
- Biofilm formation at the air–liquid interface by the Pseudomonas fluorescens SBW25 wrinkly spreader requires an acetylated form of cellulose
- Opioid Activation of Toll-Like Receptor 4 Contributes to Drug Reinforcement
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.