Impact of femoro-tibial size combinations and TKA design on kinematics
Abstract
INTRODUCTION: The variability in patients' femoral and tibial anatomy requires to use different tibia component sizes with the same femoral component size. These size combinations are allowed by manufacturers, but the clinical impact remains unclear. Therefore, the goals of our study were to investigate whether combining different sizes has an impact on the kinematics for two well-established knee systems and to compare these systems' kinematics to the native kinematics.
MATERIALS AND METHODS: Six fresh frozen knee specimens were tested in a force controlled knee rig before and after implantation of a cruciate retaining (CR) and a posterior-stabilized (PS) implant. Femoro-tibial kinematics were recorded using a ultrasonic-based motion analysis system while performing a loaded squat from 30° to 130°. In each knee, the original best fit inlay was then replaced by different inlays simulating a smaller or bigger tibia component. The kinematics obtained with the simulated sizes were compared to the original inlay kinematics using descriptive statistics.
RESULTS: For all size combinations, the difference to the original kinematics reached an average of 1.3 ± 3.3 mm in translation and - 0.1 ± 1.2° in rotation with the CR implant. With the PS implant, the average differences reached 0.4 ± 2.7 mm and - 0.2 ± 0.8°. Among all knees, no size combination consistently resulted in significantly different kinematics. Each knee showed a singular kinematic pattern. For both knee systems, the…
Read the paper
Related studies
- Combining in silico and in vitro approaches to identification of potent inhibitor against phospholipase A2 (PLA2)
- In silico and in vitro studies: investigating the chemical composition, DFT, molecular docking, and dynamic simulation of Satureja candidissima (Munby) Briq essential oil as a potential antibacterial agent
- Quercetin combined with ciprofloxacin and gentamicin inhibits biofilm formation and virulence in Staphylococcus aureus
- In silico study of the mechanism of action, pharmacokinetic and toxicological properties of some N-methylanthranilates and their analogs
- Anticonvulsant Effects of Citronellal, Daidzin, and Phytol, Possibly Through GABAA Receptor and Voltage-Gated Sodium Channel Interaction Pathways: In Vivo and In Silico Studies
- 2-Phenylcyclopropylmethylamine (PCPMA) Derivatives as D3R-Selective Ligands for 3D-QSAR, Docking and Molecular Dynamics Simulation Studies
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.