Thin slice photon-counting CT coronary angiography compared to conventional CT: Objective image quality and clinical radiation dose assessment
Abstract
BACKGROUND: Photon-counting CT (PCCT) is the next-generation CT scanner that enables improved spatial resolution and spectral imaging. For full spectral processing, higher tube voltages compared to conventional CT are necessary to achieve the required spectral separation. This generated interest in the potential influence of thin slice high tube voltage PCCT on overall image quality and consequently on radiation dose.
PURPOSE: This study first evaluated tube voltages and radiation doses applied in patients who underwent coronary CT angiography with PCCT and energy-integrating detector CT (EID-CT). Next, image quality of PCCT and EID-CT was objectively evaluated in a phantom study simulating different patient sizes at these tube voltages and radiation doses.
METHODS: We conducted a retrospective analysis of clinical doses of patients scanned on a conventional and PCCT system. Average patient water equivalent diameters for different tube voltages were extracted from the dose reports for both EID-CT and PCCT. A conical phantom made of polyethylene with multiple diameters (26/31/36 cm) representing different patient sizes and containing an iodine insert was scanned with a EID-CT scanner using tube voltages and phantom diameters that match the patient scans and characteristics. Next, phantom scans were made with PCCT at a fixed tube voltage of 120 kV and with CTDIVOL values and phantom diameters identical to the EID-CT scans. Clinical image reconstructions at 0.6 mm slice…
How strong is the evidence?
Evidence Hub reads this as a cohort study, done in people. Its evidence level is 3 of 5 (Moderate · trial / cohort / secondary analysis). Trials without a large randomized design, cohorts and secondary analyses give moderate evidence. Reported size: 143 participants or records.
Read the paper
Related studies
- Chaos and quantum thermalization
- The Dalton quantum chemistry program system
- Cord Blood Leptin and Adiponectin as Predictors of Adiposity in Children at 3 Years of Age: A Prospective Cohort Study
- Quantitative metabolomics by H-NMR and LC-MS/MS confirms altered metabolic pathways in diabetes
- Stable adiabatic quantum batteries
- A novel self-lipid antigen targets human T cells against CD1c(+) leukemias
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.