Mathematical estimation of half-value layer thicknesses

Omoumi FH, Wu X, Ghani MU, Wong MD, Li Y, Liu H

Journal of applied clinical medical physics, 2021

Abstract

OBJECTIVE: The objective of this article is to introduce a simplified and swift method to satisfactorily estimate the half-value layers (HVL), quarter-value layer (QVL), and tenth-value layer (TVL) from the x-ray spectra emitted by any diagnostic radiology or kV radiotherapy x-ray tubes.

METHODS: A CdTe x-ray and Gamma detector (X-123 CdTe, AmpTek Inc.) is used to measure the x-ray spectra at four different x-ray energies (low, mid, high energy x-rays) with different external filtering. The software "SpekCalc GUI" (Developed in McGill University, Montreal, Canada) is also used to obtain the simulated x-ray spectra. Both measured and simulated spectra are used to compute the HVL thicknesses of Aluminum by a mathematical method presented in this article. Next, the HVL thicknesses for corresponding tube potentials are also measured by calibrated ionization chamber and varying thicknesses of aluminum plates. Finally, the computed and measured HVL, QVL, and TVL thicknesses are compared to evaluate the efficacy of the presented method.

RESULTS: The results show acceptable concordance between computed and measured quantities. The disagreement rates between measured HVL and the values derived mathematically from the x-ray spectra are 10 to 90 micrometers of Aluminum at tube potentials of 31 kV to 120 kV. As it is shown, a negligible discrepancy is observed between the analytical estimation and the experimental assessments.

CONCLUSION: The HVL is an essential component in the…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free