Assessment of the spectral performance of hybrid photon counting x-ray detectors

Trueb P, Zambon P, Broennimann C

Medical physics, 2017

Abstract

PURPOSE: Hybrid Photon Counting (HPC) detectors profoundly improved x-ray diffraction experiments at third generation synchrotron facilities. Enabling the simultaneous measurement of x-ray intensities in multiple energy bins, they also have many potential applications in the field of medical imaging. A prerequisite for this is a clean spectral response. To quantify how efficiently HPC detectors are able to assign photons to the correct energy bin, a quantity called Spectral Efficiency (SE) is introduced. This figure of merit measures the number of x-rays with correctly assigned energy normalized to the number of incoming photons.

METHODS: A prototype HPC detector has been used to perform precision measurements of x-ray spectra at the BESSY synchrotron. The detector consists of a novel ASIC with pixels of 75 × 75 μm2 size and a 750 μm thick CdTe sensor. The experimental data are complemented by the results of a Monte-Carlo (MC) simulation, which not only includes the physical detection process but also pulse pile-up at high photon fluxes. The spectra and the measured photon flux are used to infer the Spectral Efficiency.

RESULTS: In the energy range from 10 to 60 keV, both the Quantum Efficiency and the Spectral Efficiency were precisely measured and simulated. Good agreement between simulation and experiment has been achieved. For the small pixels of the prototype detector, a SE between 15% and 77% has been determined. The MC simulation is used to predict the SE for various…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free