Evidence Hub

Detective quantum efficiency of an amorphous selenium detector to megavoltage radiation

Mah D, Rawlinson JA, Rowlands JA

Physics in medicine and biology, 1999

Abstract

The spatial frequency dependent detective quantum efficiency (DQE(f)) of a high-resolution selenium-based imaging system has been measured at megavoltage energies. These results have been compared with theoretical calculations. The imaging system was a video tube with a 5 microm amorphous selenium (a-Se) target which was irradiated by 1.25 MeV gamma-rays. The modulation transfer function (MTF) decreased rapidly with spatial frequency (determined by spread of electrons in the build-up material) while the noise power spectrum was constant as a function of spatial frequency. The DQE obtained from these MTF and noise power measurements was compared with a Monte Carlo model of the pulse height spectrum of the detector. The DQE(0) model accounted for the interaction of x rays with the detector as well as the energy-dependent gain (charge generated/energy deposition). Good agreement between the calculated and measured DQE(0) was found. The model was also used to estimate the DQE(f) of a metal plate + a-Se detector which was compared with a metal plate + phosphor system of the same mass thickness. The DQE(f) s of both detectors are very similar, indicating that the choice of which detector is better will be based upon criteria other than DQE(f), such as read-out approach, ease of manufacture or sensitivity.

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.