Comparison of physical approaches to inter-crystal scattering in a staggered-layer scintillation PET detector: a simulation study

Liu Q, Scheins JJ, Pfaehler E, George Ciantar K, Niekämper D, Régio Brambilla C, Zhang Q, Jon Shah N et al.

Physics in medicine and biology, 2026

Abstract

Abstract Objective. Inter-crystal scattering (ICS) degrades spatial resolution and contrast by causing spurious lines of response in small-pitch, staggered three-layer pixelated LSO scintillation detectors used in a high-resolution BrainPET-7T insert. This work aimed to investigate the distribution of ICS events in our BrainPET-7T detector and understand its underlying mechanisms. Approach. Six physical approaches for computing probabilities to identify the first interaction positions of ICS events were studied using Geant4 simulations, focusing specifically on events with two interactions, which represent the most frequent ICS occurrence. Performance was evaluated using ideal interaction positions and energies directly from simulations, ground truth crystal-center positions with ideal energies, as well as crystal-center positions and energies determined by a maximum likelihood positioning (MLP) algorithm. Main results. A method based on Klein–Nishina probabilities computed using geometric information (KN-G), together with MLP-derived crystal-center positions and energies, achieved superior accuracy, particularly when the distance between successive interactions in Compton events was shorter than 6 mm and when incident angles were below 15 ∘ . This advantage is partly attributable to crystal-center discretization, which masks a substantial portion of backscattering events. It reached a peak accuracy of 92.2% at a distance of 19 mm. Significance. These findings demonstrate…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free