Evidence Hub

Monte Carlo modeling of radiation dose from radiation therapy with superficial x-rays

Barghash R, Martin TW, Prebble AR, Leary D

Journal of applied clinical medical physics, 2025

2 citations

Abstract

Introduction: Superficial x-rays (50-100 kVp) are used for treating non-melanoma skin cancer and intraoperative radiation therapy (IORT). At these energies, the photoelectric effect significantly increases absorbed dose to bone compared to soft tissue.

Methods: We used EGSnrc MC simulations to investigate bone dose enhancement during radiotherapy with the Sensus SRT-100 machine. Simulated beams were validated against laboratory measurements and compared to a commercial treatment planning system (SmART-ATP). Transmission simulations indirectly predicted bone dosage. Simulated beams were utilized as a mock treatment plan from a human cone-beam computed tomography (CBCT) image.

Results: EGSnrc accurately modeled the Sensus SRT-100 beams (100, 70, and 50 kVp) with a root mean square error (RMSE) of percentage depth dose ratios between Monte Carlo predictions and lab measurements of 1.66, 0.47, and 0.99, respectively. PDDs from simulations of a water phantom with bone slabs showed peak doses at water-bone interfaces relative to surface doses. At 0.3 cm depth bone slab, doses reached 410%, 490%, and 510% for 50, 70, and 100 kVp, respectively. At 1.5 cm depth, doses were 140%, 215%, and 270%. At 2.5 cm depth, peak doses were 74%, 130%, and 170% for 50, 70, and 100 kVp, respectively. A simulated treatment plan (4 Gy surface dose) using a CBCT of a human head predicted the dose to the skull to be around 20, 19, and 15 Gy for the 100, 70, and 50 kVp beams, respectively.

Conclusions:…

How strong is the evidence?

Evidence Hub reads this as a simulation / modelling, done in computer models (in silico), not in people. Its evidence level is 5 of 5 (Very low · case / narrative / animal / in vitro / in silico). Case reports, narrative reviews, and animal, cell or computer-model studies give the weakest evidence for human health.

How the five levels of evidence work. The level is assigned automatically from the abstract and can be wrong, so check the paper.

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free