Numerical evaluation of high-energy, laser-Compton x-ray sources for contrast enhancement and dose reduction in clinical imaging via gadolinium-based K-edge subtraction
Abstract
Conventional x-ray sources for medical imaging utilize bremsstrahlung radiation. These sources generate large bandwidth (BW) x-ray spectra with large fractions of photons that impart a dose, but do not contribute to image production. X-ray sources based on laser-Compton scattering can have inherently small energy BWs and can be tuned to low dose-imparting energies, allowing them to take advantage of atomic K-edge contrast enhancement. This paper investigates the use of gadolinium-based K-edge subtraction imaging in the context of mammography using a laser-Compton source through simulations quantifying contrast and dose in such imaging systems as a function of laser-Compton source parameters. Our simulations indicate that a K-edge subtraction image generated with a 0.5% BW (FWHM) laser-Compton x-ray source can obtain an equal contrast to a bremsstrahlung image with only 3% of the dose.
Read the paper
Related studies
- Characterization and Analysis of Porosity and Pore Structures
- Practical considerations for conducting ecotoxicity test methods with manufactured nanomaterials: what have we learnt so far?
- Dental Optical Coherence Tomography
- Radiation-shielding concrete: A review of materials, performance, and the impact of radiation on concrete properties
- Testing entanglement of annihilation photons
- Calanus finmarchicus seasonal cycle and diapause in relation to gene expression, physiology, and endogenous clocks
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.