Evidence Hub

Performance evaluation of MACACO: a multilayer Compton camera

E. Muñoz, J. Barrio, A. Etxebeste, P. G. Ortega, C. Lacasta, J. Oliver, C. Solaz, G. Llosá

Physics in Medicine and Biology, 2017

59 citations

Abstract

Compton imaging devices have been proposed and studied for a wide range of applications. We have developed a Compton camera prototype which can be operated with two or three detector layers based on monolithic lanthanum bromide ([Formula: see text]) crystals coupled to silicon photomultipliers (SiPMs), to be used for proton range verification in hadron therapy. In this work, we present the results obtained with our prototype in laboratory tests with radioactive sources and in simulation studies. Images of a [Formula: see text]Na and an [Formula: see text]Y radioactive sources have been successfully reconstructed. The full width half maximum of the reconstructed images is below 4 mm for a [Formula: see text]Na source at a distance of 5 cm.

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.