Compton coincidence in silicon photon-counting CT detectors

Christel Sundberg, M. Danielsson, M. Persson

Journal of Medical Imaging, 2021

7 citations

Abstract

Abstract. Purpose: Compton interactions amount to a significant fraction of the registered counts in a silicon detector. In a Compton interaction, only a part of the photon energy is deposited and a single incident photon can result in multiple counts unless tungsten shielding is used. Deep silicon has proved to be a competitive material for photon-counting CT detectors, but to improve the performance further, one possibility is to use coincidence techniques to identify Compton-scattered photons and reconstruct their incident energies. Approach: In a detector with no tungsten shielding, incident photons can interact through a series of interactions. Based on the position and energy of each interaction, probability-based methods can be used to estimate the incident photon energy. Here, we present a maximum likelihood estimation framework along with an alternative method to estimate the incident photon energy and position in a silicon detector. Results: Assuming one incident photon per time frame, we show that the incident photon energy can be estimated with a mean error of −0.07 ± 0.03 keV and an RMS error of 3.36 ± 0.02 keV for a realistic case in which we assume a detector with limited energy and spatial resolution. The interaction position was estimated with a mean error of −2 ± 11 μm in x direction and 7 ± 11 μm in y direction. Corresponding RMS errors of 1.09 ± 0.01 and 1.10 ± 0.01 mm were achieved in x and y, respectively. Conclusions: The presented results show the…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free