Dosimetric characteristics of brass mesh as bolus under megavoltage photon irradiation
Abstract
Objective: This article presents a set of dosimetric measurements describing the properties of brass mesh (Whiting and Davis, Attleboro Falls, MA) under megavoltage photon irradiation conditions, with particular relevance to its use in breast radiotherapy.
Methods: The effectiveness of brass mesh as a bolus material was investigated using 6-, 15- and 6-MV flattening filter-free photon beams. The effect on dose build-up at the entrance surface, build-down at the beam-exit surface, dose with surface entrance obliquity, beam profiles, penumbra and percentage depth doses were investigated.
Results: One layer of the brass mesh produces a build-up effect equivalent to 1.1 mm of water at 6 MV and 1.9 mm at 15 MV. The brass generates a backscattered component of dose, if the photon beam exits through it. Percentage depth-dose curves are largely unaffected by the mesh and are shown to be equivalent to plain-field data. Beam penumbra and profiles are unchanged by the brass except within the first millimetre (mm) of phantom, where a periodic pattern of dose enhancement is seen.
Conclusion: The data presented demonstrate that one layer of brass mesh provides a similar dose build-up effect equivalent to only a few millimetres of water. However, backscatter from the high atomic number (Z) mesh, at the beam exit, contributes appreciably to the overall dose surface enhancement. This dosimetric consequence cannot be neglected and indeed should be considered and accounted for, when…
Read the paper
Related studies
- Characterization and Analysis of Porosity and Pore Structures
- Practical considerations for conducting ecotoxicity test methods with manufactured nanomaterials: what have we learnt so far?
- Dental Optical Coherence Tomography
- Radiation-shielding concrete: A review of materials, performance, and the impact of radiation on concrete properties
- Testing entanglement of annihilation photons
- Calanus finmarchicus seasonal cycle and diapause in relation to gene expression, physiology, and endogenous clocks
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.