Effect of inhomogeneities and source position on dose distribution of nucletron high dose rate Ir-192 brachytherapy source by Monte Carlo simulation

Rakesh Mohan Chandola, Samit Tiwari, MK Kowar, Vivek Choudhary

Journal of Cancer Research and Therapeutics, 2010

23 citations

Abstract

BACKGROUND: The presence of least dense dry air and highly dense cortical bone in the path of radiation and the position of source, near or far from the surface of patient, affects the exact dose delivery like in breast brachytherapy. AIM: This study aims to find out the dose difference in the presence of inhomogeneities like cortical bone and dry air as well as to find out difference of dose due to position of source in water phantom of high dose rate (HDR) 192 Ir nucletron microselectron v2 (mHDRv2) brachytherapy source using Monte Carlo (MC) simulation EGSnrc code, so that the results could be used in Treatment Planning System (TPS) for more precise brachytherapy treatment. SETTINGS AND DESIGN: The settings and design are done using different software of the computer. METHODS AND MATERIALS: For this study, the said source, water phantom of volume 30 x 30 x 30 cm 3, inhomogeneities each of volume 1 x 2 x 2 cm3 with their position, water of water phantom and position of source are modeled using three-dimensional MC EGSnrc code. STATISTICAL ANALYSIS USED: Mean and probability are used for results and discussion. RESULTS: The % relative dose difference is calculated here as 5.5 to 6.5% higher and 4.5 to 5% lower in the presence of air and cortical bone respectively at transverse axis of the source, which may be due to difference of linear attenuation coefficients of the inhomogeneities. However, when the source was positioned at 1 cm distance from the surface of water…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free