Dosimetry for synchrotron x-ray microbeam radiation therapy
Abstract
In this work different dosimetric quantities of relevance for a new kind of radiation therapy, x-ray microbeam radiation therapy, have been both calculated and measured. A detailed analysis of the microbeam dose deposition produced by a single microbeam and by an array of microbeams has been performed using Monte Carlo simulations of the radiation transport. The peak-to-valley dose ratio was found to be decreasing with depth in the irradiated object. The dose in air (or more correctly the air kerma) produced by the high-flux synchrotron-generated x-ray beam was measured with a thimble-type ionization chamber in the uncollimated beam using two different approaches. The x-ray spectrum was measured with a powder diffraction method. The radiochromic film was found useful for rapid qualitative measurements of the dose distributions produced by the microbeams. The dose inside and near an x-ray microbeam was also measured with a MOSFET detector.
Read the paper
Related studies
- Characterization and Analysis of Porosity and Pore Structures
- Practical considerations for conducting ecotoxicity test methods with manufactured nanomaterials: what have we learnt so far?
- Dental Optical Coherence Tomography
- Radiation-shielding concrete: A review of materials, performance, and the impact of radiation on concrete properties
- Testing entanglement of annihilation photons
- Calanus finmarchicus seasonal cycle and diapause in relation to gene expression, physiology, and endogenous clocks
Go deeper in Evidence Hub