PURPOSE: To describe an animal model used to evaluate the propensity of various biomaterials to calcify intraocularly. SETTING: Research Department, Allergan Inc., Irvine, California, USA. METHODS: Intraocular lens (IOL) optic materials were implanted intramuscularly and/or subcutaneously in rabbits for up to 90 days. The materials included silicone, poly(methyl methacrylate) (PMMA), hydroxyethyl methacrylate hydrogel, and several hydrophobic acrylic materials. Scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDS) were used to detect calcification demonstrated by characteristic discrete nodules containing both calcium and phosphate. Histological methods were used to evaluate tissue reactivity. Disc lenses fabricated from the experimental material were also bilaterally implanted in rabbit eyes that were monitored by slitlamp biomicroscopy. The lenses were explanted at 1, 2, 5.5, 10, and 20 months for SEM/EDS analysis. RESULTS: No calcification was noted in the intramuscularly or subcutaneously implanted silicone, PMMA, and acrylic optic materials. Calcification was noted on the intramuscularly, subcutaneously, and intraocularly implanted experimental acrylic and the intramuscularly implanted hydrogel material; the calcification was more extensive on the hydrogel. Signs that suggested intraocular calcification were first noted on the experimental IOLs at 4 months, but calcification was not confirmed until 10 months. CONCLUSIONS: Material…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free