Characteristics of iodine-123 IQ-SPECT/CT imaging compared with conventional SPECT/CT.

Shibutani T, Onoguchi M, Yoneyama H, Konishi T, Matsuo S, Nakajima K

Annals of nuclear medicine, 2019

Abstract

OBJECTIVES: Although the utility of IQ-SPECT imaging using 99mTc and 201Tl myocardial perfusion SPECT has been reported, 123I-labeled myocardial SPECT has not been fully evaluated. We determined the characteristics and utility of 123I IQ-SPECT imaging compared with conventional SPECT (C-SPECT).

METHODS: Two myocardial phantom patterns were used to simulate normal myocardium and myocardial infarction. SPECT acquisition was performed using a hybrid dual-head SPECT/CT system equipped with a SMARTZOOM collimator for IQ-SPECT or a low-medium energy general purpose collimator for C-SPECT. Projection data were reconstructed using ordered subset expectation maximization with depth-dependent 3-dimensional resolution recovery for C-SPECT and ordered subset conjugate gradient minimizer method for IQ-SPECT. Three types of myocardial image were created; namely, no correction (NC), with attenuation correction (AC), and with both attenuation and scatter corrections (ACSC). Five observers visually scored the homogeneity of normal myocardium and defect severity of the myocardium with inferior defects by a five-point scale: homogeneity scores (5 = homogeneous to 1 = inhomogeneous) and defect scores (5 = excellent to 1 = poor). We also created a 17-segment polar map and quantitatively assessed segmental %uptake using a myocardial phantom with normal findings and defects.

RESULTS: The average visual homogeneity scores of the IQ-SPECT with NC and ACSC were significantly higher than that of…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free