Evidence Hub

Low-dose megavoltage cone-beam computed tomography using a novel multi-layer imager (MLI)

Myronakis M, Huber P, Lehmann M, Fueglistaller R, Jacobson M, Hu YH, Baturin P, Wang A et al.

Medical physics, 2020

Abstract

PURPOSE: The feasibility of low-dose megavoltage cone-beam acquisition (MVCBCT) using a novel, high detective quantum efficiency (DQE) multi-layer imager (MLI) was investigated. The aim of this work was to reconstruct MVCBCT images using the MLI at different total dose levels, and assess Hounsfield Unit (HU) accuracy, noise and contrast-to-noise ratio (CNR) for low-dose megavoltage cone-beam acquisition.

METHODS: The MLI has four stacked layers; each layer contains a combination of copper filter/converter, gadolinium oxysulfide (GOS) scintillator and a-Si detector array. In total, 720 projections of a CATPHAN® phantom were acquired over 360° at 2.5, 6, and 6 MV flattening filter free (FFF) beam energies on a Varian TrueBeam LINAC. The dose per projection was 0.01, 0.0167, and 0.05 MU for 2.5, 6, and 6 MV FFF, respectively. MVCBCT images were reconstructed with varying numbers of projections to provide a range of doses for evaluation. Hounsfield Unit uniformity, accuracy, noise and CNR were estimated. Improvements were quantified relative to the standard AS1200 single-layer imager.

RESULTS: Average HU uniformity for the MLI reconstructions was within a range of 95%-99% for all of the energies studied. Relative electron density estimation from HU values was within 0.4% ± 1.8% from nominal values. The CNR for MVCBCT based on MLI projections was 2-4× greater than from AS1200 projections. The 2.5 MV beam acquisition with the MLI exhibited the lowest noise and the best balance…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free