Portable Single-Exposure Dual-Energy X-ray Detector for Improved Point-of-Care Diagnostic Imaging

Karim KS, Tilley Ii S

Military medicine, 2023

Abstract

INTRODUCTION: Dual-energy subtraction (DES) imaging is well known to reduce anatomical noise and enable material classification. The current approaches to DES imaging have trade-offs, such as motion artifacts, low sensitivity because of losses in a mid-filter, and lack of portability. Recently, a portable triple-layer flat-panel detector (FPD) was proposed for use in single-shot DES imaging that can provide improved sensitivity and removal of motion artifacts in a point-of-care setting. The purpose of this study is to evaluate the feasibility of such a detector. Various image quality metrics and clinical images are provided.

MATERIALS AND METHODS: An FDA-cleared single-exposure DES FPD consisting of three stacked sensors was used for all measurements. This detector generates three images per exposure: A digital radiography (DR) image, i.e., as would be produced with a conventional detector, and two DES images, bone and soft tissue. To evaluate DR image quality, detective quantum efficiency (DQE) and modulation transfer function were measured for multiple radiation quality beams. Digital radiography and DES images obtained from this FPD were evaluated in previously reported fixed and portable clinical studies. Digital radiography and DES images from case studies are presented for qualitative assessment.

RESULTS: Modulation transfer function and DQE were measured across multiple radiation quality beams for the DR image. The DES images showed good tissue separation and…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free