Occupational and patient exposure as well as image quality for full spine examinations with the EOS imaging system
Abstract
PURPOSE: EOS (EOS imaging S.A, Paris, France) is an x-ray imaging system that uses slot-scanning technology in order to optimize the trade-off between image quality and dose. The goal of this study was to characterize the EOS system in terms of occupational exposure, organ doses to patients as well as image quality for full spine examinations.
METHODS: Occupational exposure was determined by measuring the ambient dose equivalents in the radiological room during a standard full spine examination. The patient dosimetry was performed using anthropomorphic phantoms representing an adolescent and a five-year-old child. The organ doses were measured with thermoluminescent detectors and then used to calculate effective doses. Patient exposure with EOS was then compared to dose levels reported for conventional radiological systems. Image quality was assessed in terms of spatial resolution and different noise contributions to evaluate the detector's performances of the system. The spatial-frequency signal transfer efficiency of the imaging system was quantified by the detective quantum efficiency (DQE).
RESULTS: The use of a protective apron when the medical staff or parents have to stand near to the cubicle in the radiological room is recommended. The estimated effective dose to patients undergoing a full spine examination with the EOS system was 290 μSv for an adult and 200 μSv for a child. MTF and NPS are nonisotropic, with higher values in the scanning direction; they are in…
Read the paper
Related studies
- Developing high safety Li-metal anodes for future high-energy Li-metal batteries: strategies and perspectives
- Structure and Energy Transfer in Photosystems of Oxygenic Photosynthesis
- Ground- and Space-Based Gamma-Ray Astronomy
- Energy Functions in De Novo Protein Design: Current Challenges and Future Prospects
- Electronic structure of ions and molecules in solution: a view from modern soft X-ray spectroscopies
- Arising 2D Perovskites for Ionizing Radiation Detection
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.