Triple-sensitivity high-spatial-resolution X-ray computed tomography using a cadmium-telluride detector and its beam-hardening effect

Yoshida S, Sato E, Oda Y, Yoshioka K, Moriyama H, Watanabe M

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine, 2020

Abstract

To observe blood vessels at high contrasts, we constructed a first-generation triple-sensitivity X-ray computed tomography (TS-CT) scanner using a cadmium-telluride (CdTe) detector and a triple-amplifying system. X-ray photons are absorbed by the CdTe crystal, and the electric charges produced by photons are converted into voltages using a current-to-voltage (I-V) amplifier, and the I-V output is amplified by a voltage-to-voltage (V-V) amplifier. The V-V output 1 is sent to a dual V-V amplifier through a 5.0-m-length coaxial cable and amplified to two-different outputs of 2 and 3. The three outputs 1-3 are sent to a personal computer through an analog-to-digital converter to reconstruct three tomograms simultaneously. In the TS-CT, the scattering photons from the object are extremely reduced using a 0.5-mm-diameter lead pinhole behind the object. The translation and rotation steps were 0.1 mm and 0.5°, respectively, and the spatial resolutions were 0.25 × 0.25 mm2. The scanning time was 19.6 min, and blood vessels were visible using gadolinium contrast media. In particular, the effective photon energy increased with increasing amplification factor of the amplifier caused by beam hardening of the object.

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free