Technical note: Exploring the detectability of coronary calcification using ultra-high-resolution photon-counting-detector CT.

Chang S, Ren L, Tang S, Marsh JF, Hsieh S, McCollough CH, Leng S

Medical physics, 2023

Abstract

BACKGROUND: Coronary calcification is a strong indicator of coronary artery disease, and patients with a "zero" coronary calcification score have a much lower risk of future cardiac events than those with even small amounts of calcium. However, false-negative (incorrect zero scores) may occur if small calcifications are missed at CT due to limited spatial resolution.

PURPOSE: To demonstrate lower limits of detection for coronary calcification using an ultra-high-resolution (UHR) mode on a clinical photon-counting-detector CT (PCD-CT), compared to a conventional energy-integrating-detector CT (EID-CT).

METHODS: Chicken eggshell fragments (0.4-0.8 mm) mimicking coronary calcifications were scanned on a clinical PCD-CT (NAEOTOM Alpha) in UHR mode and a conventional EID-CT (SOMATOM Force) with matched tube potential and radiation dose levels to the PCD-CT. PCD-CT images were reconstructed with a sharp kernel (Qr68) and a quantum iterative algorithm (QIR-3). Two sets of EID-CT images were reconstructed: routine clinical kernel (Qr36, ADMIRE-3) and a sharper kernel (Qr54) with similar noise to PCD-CT images. With institutional review board approval, in vivo exams performed with the PCD-CT in UHR mode were compared against patients' clinical EID-CT exams. The visibility of calcifications on PCD-CT and EID-CT images was assessed and compared qualitatively.

RESULTS: PCD-CT images visualized all calcified fragments, while EID-CT failed to detect those below 0.6 mm using a routine…