A head phantom study for intraocular dose evaluation of 64-slice multidetector CT examination in patients with suspected cranial trauma

Matsubara K, Koshida K, Noto K, Takata T, Suzuki M, Shimono T, Yamamoto T, Matsui O

European journal of radiology, 2011

Abstract

PURPOSE: In cases of suspected cranial trauma, cranial CT examinations should be performed to rule out pathology. There are some methods available for reducing intraocular doses; however, it is difficult for the operators to conduct the necessary measurements because of restrictions in time and patient mobility, especially in high-energy trauma cases. Therefore, we performed a head phantom study for intraocular dose evaluation of 64-slice multidetector CT examination in patients with suspected cranial trauma.

MATERIALS AND METHODS: Assuming that the orbitomeatal (OM) line and bed were vertical, a head phantom was tilted from 10 degrees caudally to 25 degrees cranially at 5-degree intervals. At each tilted position, the phantom was examined using a 64-section multidetector CT device using three acquisition protocols. Intraocular doses during each examination were measured using small dosimeters.

RESULTS: Assuming that the OM line and bed were vertical, intraocular doses varied between 52 and 140%, 17-138%, and 90-142% during helical, non-helical, and helical CT angiographic examinations, respectively. Intraocular doses increased when the phantom was tilted cranially.

CONCLUSION: If possible, the best way to reduce the intraocular dose is by angling the gantry cranially, tilting the head of each patient caudally and adopting a non-helical acquisition method. During procedure, the acquisition angle should be angled cranially more than 0 degrees based on the OM line. The…