Long-length tomosynthesis and 3D-2D registration for intraoperative assessment of spine instrumentation

Xiaoxuan Zhang, A. Uneri, Pengwei Wu, M. Ketcha, Craig K. Jones, Yixuan Huang, S. Lo, P. Helm et al.

Physics in Medicine and Biology, 2021

16 citations

Abstract

Purpose. A system for long-length intraoperative imaging is reported based on longitudinal motion of an O-arm gantry featuring a multi-slot collimator. We assess the utility of long-length tomosynthesis and the geometric accuracy of 3D image registration for surgical guidance and evaluation of long spinal constructs. Methods. A multi-slot collimator with tilted apertures was integrated into an O-arm system for long-length imaging. The multi-slot projective geometry leads to slight view disparity in both long-length projection images (referred to as ‘line scans’) and tomosynthesis ‘slot reconstructions’ produced using a weighted-backprojection method. The radiation dose for long-length imaging was measured, and the utility of long-length, intraoperative tomosynthesis was evaluated in phantom and cadaver studies. Leveraging the depth resolution provided by parallax views, an algorithm for 3D-2D registration of the patient and surgical devices was adapted for registration with line scans and slot reconstructions. Registration performance using single-plane or dual-plane long-length images was evaluated and compared to registration accuracy achieved using standard dual-plane radiographs. Results. Longitudinal coverage of ∼50–64 cm was achieved with a single long-length slot scan, providing a field-of-view (FOV) up to (40 × 64) cm2, depending on patient positioning. The dose-area product (reference point air kerma × x-ray field area) for a slot scan ranged from ∼702–1757 mGy·cm2,…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free