Quantitative accuracy of 177Lu SPECT reconstruction using different compensation methods: phantom and patient studies

Eero Hippeläinen, M. Tenhunen, H. Mäenpää, A. Sohlberg

EJNMMI Research, 2016

50 citations

Abstract

BackgroundIn targeted radionuclide therapy (TRT), accurate quantification using SPECT/CT images is important for optimizing radiation dose delivered to both the tumour and healthy tissue. Quantitative SPECT images are regularly reconstructed using the ordered subset expectation maximization (OSEM) algorithm with various compensation methods such as attenuation (A), scatter (S) and detector and collimator response (R). In this study, different combinations of the compensation methods are applied during OSEM reconstruction and the effect on the 177Lu quantification accuracy is studied in an anthropomorphic torso phantom. In addition, the phantom results are reflected to (177)Lu-DOTA-Tyr3-octreotate (177Lu-DOTATATE)-treated patient data and kidney absorbed dose estimates.MethodsThe torso phantom was imaged with nine various sized (0.4–104.4 cm3) spherical inserts, filled with known 177Lu activity ranging from 0.5 to 105.5 MBq. Images were reconstructed using OSEM algorithm using A, AR and ARS compensation method combinations. The compensation method combinations were compared by calculating the concentration recovery coefficient (cRC) for each insert. In addition, ten 177Lu-DOTATATE-treated patient’s post-therapy dosimetry acquisitions were reconstructed, and the absorbed dose to kidneys was estimated.ResultscRC values depend on the insert size for all compensation methods. AR and ARS produced significantly higher cRC values than attenuation correction alone. There were no cRC…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free