Evidence Hub

Comprehensive stereotactic radiosurgery platform characterization: A novel end-to-end approach with anthropomorphic 3D dosimetry

Yi-Fang Wang, J. Adamovics, C. Wuu

Medical Physics (Lancaster), 2024

1 citation

Abstract

BACKGROUND

Stereotactic radiosurgery (SRS) is a widely employed strategy for intracranial metastases, utilizing linear accelerators and volumetric modulated arc therapy (VMAT). Ensuring precise linear accelerator performance is crucial, given the small planning target volume (PTV) margins. Rapid dose falloff is vital to minimize brain radiation necrosis. Despite advances in SRS planning, tools for end-to-end testing of SRS treatments are lacking, hindering confidence in the procedure.

PURPOSE

This study introduces a novel end-to-end three-dimensional (3D) anthropomorphic dosimetry system for characterization of a radiosurgery platform, aiming to measure planning metrics, dose gradient index (DGI), brain volumes receiving at least 10 and 12 Gy (V10, V12), as well as assess delivery uncertainties in multitarget treatments. The study also compares metrics from benchmark plans to enhance understanding and confidence in SRS treatments.

METHODS

The developed anthropomorphic 3D dosimetry system includes a modified Stereotactic End-to-End Verification (STEEV) phantom with a customized insert integrating 3D dosimeters and a fiber optic CT scanner. Labview and MATLAB programs handle optical scanning, image preprocessing, and dosimetric analysis. SlicerRT is used for 3D dose visualization and analysis. A film stack insert was used to validate the 3D dosimeter measurements at specific slices. Benchmark plans were developed and measured to investigate off-axis errors, dose spillage,…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free