Impact of bowtie filter and detector collimation on multislice CT scatter profiles: a simulation study
Abstract
PURPOSE
To investigate via Monte Carlo simulations, the impact of scan subject size, antiscatter grid (ASG), collimator size, and bow-tie filter on the distribution of scatter radiation in a typical realistically modeled third generation 16 slice diagnostic CT scanner.
METHODS
Full radiation transport was simulated with Geant4 in a realistic CT scanner geometric model, including the imaging phantom, bowtie filter (BTF), collimators and detector assembly, except for the ASGs. An analytical method was employed to quantify the probable transmission through the ASG of each photon intersecting the detector array. Normalized scatter profiles (NSP) and scatter-to-primary-ratio (SPR) profiles were simulated for 90 kVp and 140 kVp beams for different size phantoms and slice thicknesses. The impact of CT scatter on the reconstructed attenuation coefficient factor was also studied as were the modulating effects of phantom- and patient-tissue heterogeneities on scatter profiles. A method to characterize the relative spatial frequency content of sinogram signals was developed to assess the latter.
RESULTS
For the 21.4 cm diameter phantom, NSP and SPR increase linearly with collimator opening for both tube potentials, with the 90 kVp scan exhibiting slightly larger NSP and SPR. The BTF modestly modulates scatter under the phantom center, reducing the prominent off-axis lobes by factors of 1.1-1.3. The ASG reduces scatter on the central axis NSP threefold, and reduces scatter at the…
Read the paper
Related studies
- Quantitative accuracy of 177Lu SPECT reconstruction using different compensation methods: phantom and patient studies
- Technique alternatives for breast radiation oncology: Conventional radiation therapy to tomotherapy
- Acceptance procedure for the linear accelerator component of the 1.5 T MRI‐linac
- Preclinical Single Photon Emission Computed Tomography of Alpha Particle-Emitting Radium-223
- Preclinical Modeling of Image-Guided Craniospinal Irradiation for Very-High-Risk Medulloblastoma
- Hand-held dental X-ray device: Attention to correct use
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.