The Effect of Parallel-hole Collimator Material on Image and Functional Parameters in SPECT Imaging: A SIMIND Monte Carlo Study

A. Azarm, J. Islamian, B. Mahmoudian, E. Gharepapagh

World Journal of Nuclear Medicine, 2015

11 citations

Abstract

The collimator in single-photon emission computed tomography (SPECT) is a critical component of the imaging system and plays an impressive role in the imaging quality. In this study, the effect of the collimator material on the radioisotopic image and its functional parameters was studied. The simulating medical imaging nuclear detectors (SIMIND) Monte Carlo program was used to simulate a Siemens E.CAM SPECT (Siemens Medical Solutions, Erlangen, Germany) system equipped with a low-energy high-resolution (LEHR) collimator. The simulation and experimental data from the SPECT imaging modality using 99m Tc were obtained on a point source and Jaszczak phantom. Seventeen high atomic number materials were considered as LEHR collimator materials. In order to determine the effect of the collimator material on the image and functional parameters, the energy resolution, spatial resolution, contrast, and collimator characteristics parameters such as septal penetration and scatter-to-primary ratio were investigated. Energy spectra profiles, full width at half maximums (FWHMs) (mm) of the point spread function (PSF) curves, system sensitivity, and contrast of cold spheres of the Jaszczak phantom for the simulated and experiment systems have acceptability superimposed. The results of FWHM and energy resolution for the 17 collimators showed that the collimator made of 98% lead and 2% antimony could provide the best FWHM and energy resolution, 7.68 mm and 9.87%, respectively. The LEHR…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free