Dopamine Transporter SPECT with 12-Minute Scan Duration Using Multiple-Pinhole Collimators
Abstract
Visual Abstract This study evaluated the potential to reduce the scan duration in dopamine transporter (DAT) SPECT when using a second-generation multiple-pinhole (MPH) collimator designed for brain SPECT with improved count sensitivity and improved spatial resolution compared with parallel-hole and fanbeam collimators. Methods: The retrospective study included 640 consecutive clinical DAT SPECT studies that had been acquired in list mode with a triple-head SPECT system with MPH collimators and a 30-min net scan duration after injection of 181 ± 10 MBq of [123I]FP-CIT. Raw data corresponding to scan durations of 20, 15, 12, 8, 6, and 4 min were obtained by restricting the events to a proportionally reduced time interval of the list-mode data for each projection angle. SPECT images were reconstructed iteratively with the same parameter settings irrespective of scan duration. The resulting 5,120 SPECT images were assessed for a neurodegeneration-typical reduction in striatal signal by visual assessment, conventional specific binding ratio analysis, and a deep convolutional neural network trained on 30-min scans. Results: Regarding visual interpretation, image quality was considered diagnostic for all 640 patients down to a 12-min scan duration. The proportion of discrepant visual interpretations between 30 and 12 min (1.2%) was not larger than the proportion of discrepant visual interpretations between 2 reading sessions of the same reader at a 30-min scan duration (1.5%).…
Read the paper
Related studies
- Description and dosimetric verification of the PEREGRINE Monte Carlo dose calculation system for photon beams incident on a water phantom
- Fast Monte Carlo based joint iterative reconstruction for simultaneous 99mTc/ 123I SPECT imaging
- Dual modality intravascular catheter system combining pulse-sampling fluorescence lifetime imaging and polarization-sensitive optical coherence tomography
- mmWaveCam: Human Vital Motion Imaging and Focused Vital Signs’ Sensing
- Task-based Design of a Synthetic-collimator SPECT System used for Small-animal Imaging
- Multicolor fluorescence microscopy for surgical guidance using a chip-scale imager with a low-NA fiber optic plate and a multi-bandpass interference filter
Go deeper in Evidence Hub
Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.