Detection of Keratoconus With a New Biomechanical Index

Riccardo Vinciguerra, Renato Ambrósio, A. Elsheikh, C. Roberts, Bernardo T. Lopes, E. Morenghi, C. Azzolini, P. Vinciguerra

Journal of refractive surgery, 2016

437 citations

Abstract

PURPOSE

To evaluate the ability of a new combined biomechanical index called the Corvis Biomechanical Index (CBI) based on corneal thickness profile and deformation parameters to separate normal from keratoconic patients.

METHODS

Six hundred fifty-eight patients (329 eyes in each database) were included in this multicenter retrospective study. Patients from two clinics located on different continents were selected to test the capability of the CBI to separate healthy and keratoconic eyes in more than one ethnic group using the Corvis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany). Logistic regression was employed to determine, based on Database 1 as the development dataset, the optimal combination of parameters to accurately separate normal from keratoconic eyes. The CBI was subsequently independently validated on Database 2.

RESULTS

The CBI included several dynamic corneal response parameters: deformation amplitude ratio at 1 and 2 mm, applanation 1 velocity, standard deviation of deformation amplitude at highest concavity, Ambrósio's Relational Thickness to the horizontal profile, and a novel stiffness parameter. The receiver operating characteristic curve analysis of the training database showed an area under the curve of 0.983. With a cut-off value of 0.5, 98.2% of the cases were correctly classified with 100% specificity and 94.1% sensitivity. In the validation dataset, the same cut-off point correctly classified 98.8% of the cases with 98.4% specificity and 100%…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free