Quality Control for Digital Tomosynthesis in the ECOG-ACRIN EA1151 TMIST Trial

Aili K. Maki, G. Mawdsley, J. Mainprize, Etta D. Pisano, S. Shen, Olivier Alonzo-Proulx, M. J. Yaffe

Medical Physics (Lancaster), 2023

4 citations

Abstract

Background The Tomosynthesis Mammography Imaging Screening Trial (TMIST), EA1151 conducted by the Eastern Cooperative Oncology Group (ECOG) / American College of Radiology Imaging Network (ACRIN) is a randomized clinical trial designed to assess the effectiveness for breast cancer screening of digital breast tomosynthesis (TM) compared to digital mammography (DM). Equipment from multiple vendors is being used in the study. Purpose For the findings of the study to be valid and capture the true capacities of the two technology types, it is important that all equipment is operated within appropriate parameters with regard to image quality and dose. A harmonized QC program was established by a core physics team. Since there are over 120 trial sites, a centralized, automated QC program was chosen as the most practical design. This report presents results of the weekly QC testing program. A companion paper will review quality monitoring based on data from the headers of the patient images. Methods Study images are collected centrally after de-identification using the “TRIAD” application developed by ACR. The core physics team devised and implemented a minimal set of quality control (QC) tests to evaluate the tomosynthesis and 2D mammography systems. Weekly, monthly and annual testing is performed by the site mammography technologists with images submitted directly to the physics core. The weekly physics quality control tests are described: SDNR of a low-contrast mass object,…

Read the paper

Related studies

Go deeper in Evidence Hub

Evidence Hub adds a study data sheet (design, sample size, evidence level, key numbers), an AI summary and a chat grounded in the paper's text.

Open this study Sign up free